Minimum Volume Optimal Design Method of Hybrid MRI Magnet
2021 · IEEE Transactions on Applied Superconductivity
저자
Quanyue Liu, Sang‐Jin Lee, Kibum Choi, 한승용 (Seungyong Hahn)
연구주제
Superconducting Materials and Applications · Electric Motor Design and Analysis
NTIS 과학기술표준분류
에너지/자원만에너지·환경·원자력 확장
배정 근거: 지원 과제의 분류와 AI 키워드가 잡은 과제의 분류를 반씩(정책 연계 맵과 같은 방법).
AI 연구 키워드·방법
키워드 (OpenAlex)
minimum volume design field superposition high-temperature superconductors fringing field magnetic field homogeneity 7T MRI nonlinear programming
초록
The purpose of this paper is to present a minimum volume optimal design method of hybrid magnetic resonance imaging (MRI) magnet using field superposition. The hybrid MRI magnet made by high-temperature superconductor (HTS) and low-temperature superconductor (LTS) is the optimal approach of producing higher magnetic field, which also reduces volume than LTS. In this study, the field superposition theory combined with linear programming and nonlinear optimization is introduced in advance to get minimum volume. Considering the different advantages of LTS magnets and HTS magnets, three hybrid MRI magnet configurations are introduced: (1) HTS insert and LTS outsert: (2) Minimum HTS magnet under LTS magnet: (3) Minimum LTS magnet under HTS magnet. In addition, the homogeneity over the central sphere volume and the scope of 5 Gauss fringe field are considered in actively shielding MRI design. …
이 논문을 지원한 국가R&D 과제 (NTIS 성과)
| 연도 | 과제 | 부처 | 사업 | 수행기관 | NTIS 분류(1순위) |
|---|---|---|---|---|---|
| 2021 | 정보기술 미래인재 교육연구단 | 교육부 | 4단계두뇌한국21사업(R&D) | 서울대학 | – |
| 2021 | 무절연 권선법에 기반한 고에너지밀도 고온 초전도 자석핵심 요소기술 개발 | 과학기술정보통신부 | 개인기초연구(과기정통부)(R&D) | 서울대학 | 전기/전자 > 중전기기 > 초전도 기술/제품에너지 외 |
인접 그래프
노드 hover = 이웃 강조 · 1회 클릭 = 고정 · 고정된 노드 다시 클릭 = 페이지 이동
이 문서와 연결된 페이지
교수 1
연구주제 1
Superconducting Materials and Applications