Potential Joint Damage in a REBCO MRI Magnet Due to the Gradient Coil-Induced Vibration
2025 · IEEE Transactions on Applied Superconductivity
저자
Wonju Jung, Seyong Choi, Hyunsoo Park, Geonyoung Kim, Jeonghwan Park, Dongwoo Lee, Jung Tae Lee, 한승용 (Seungyong Hahn)
연구주제
Superconducting Materials and Applications · Particle accelerators and beam dynamics · Electric Motor Design and Analysis
NTIS 과학기술표준분류
에너지/자원만에너지·환경·원자력 확장
지원 과제·AI 키워드로 잡을 근거가 없어 분류하지 않았다.
AI 연구 키워드·방법
gradient coil MRI 자석 mechanical vibration 초전도 접합
키워드 (OpenAlex)
joint destruction mechanical resonance natural frequencies pancake coils joint stress REBCO tapes joint space width
초록
This study investigates the potential for joint damage in REBCO MRI magnets due to mechanical vibration induced by gradient coil operation. A 6 T REBCO MRI magnet, developed as part of the Korea Medical Device Development Fund (KMDF) project, was analyzed to assess its susceptibility to gradient coil-induced vibrations. The eigenfrequency range of the magnet was identified to be within 1,787–2,000 Hz, which overlaps with the operational frequency range of the gradient coil (1–2,000 Hz). Simulations using a simplified two single pancake (SP) coil model revealed that peak stress levels at the joints could reach up to 700 MPa during resonance, approaching the tensile limit of REBCO tapes. The study suggests operating the gradient coil at frequencies at least 500 Hz away from the identified resonance range to prevent potential joint damage. Additionally, the use of wider joints between SP co…
이 논문을 지원한 국가R&D 과제 (NTIS 성과)
NTIS 성과DB(2020–24)에서 DOI 매칭 없음
인접 그래프
노드 hover = 이웃 강조 · 1회 클릭 = 고정 · 고정된 노드 다시 클릭 = 페이지 이동
이 문서와 연결된 페이지
교수 1
연구주제 1
Superconducting Materials and Applications